The S700 modular system is very easy to configure to provide a tailor-made application. An S700 housing can be equipped with up to 3 analyzer modules for compact and cost-effective system solutions. A total of 6 different analyzer modules are available for analyzing more than 60 gas components.
Depending on the measuring task, installation site, and ambient conditions, the following housing types are available:
- S710 19″ rack for analyzer cabinets and standard applications
- S715 wall-mounting enclosure for harsh application conditions and zone 2 explosive environment (ATEX)
- S720 flame-proof housing for zone 1 explosive environment (ATEX)
Specifications
Measuring principle
NDIR
Ambient temperatures
+5 °C ... +45 °C
Interfaces
RS-232c
Bus protocol
Modbus RTU
EX approval
Yes as option
Supply voltage
100 V / 115 V / 230 V - 48 ... 62 Hz
Analogue outputs
4x 4 ... 20mA, max. load 500Ω;
0 ... 10 V, electrically isolated
0 ... 10 V, electrically isolated
Analogue inputs
2x 0/2/4 ... 20mA
Option: 0 ... 10V DC
Option: 0 ... 10V DC
Digital outputs
8 Relay contacts:
30 V AC, 500mA / 48V DC, 500mA
3 relay outputs preset for failure, service and maintenance
8 Transistor outputs:
24V DC, 500mA Freely adjustable
30 V AC, 500mA / 48V DC, 500mA
3 relay outputs preset for failure, service and maintenance
8 Transistor outputs:
24V DC, 500mA Freely adjustable
Digital inputs
8 Optical coupler inputs:
24V DC electrically isolated, freely programmable
24V DC electrically isolated, freely programmable
Detection limit
< 0,5% from the range
EN14181 QAL1 certificate
Yes
EN 15267-3 certificate
Yes
Heat up time
45 minuten
Measuring gas temperature
0 °C ... +45 °C
Temperatuur bij input analysator
Temperatuur bij input analysator
Menu languages
Duits, Engels, Frans, Italiaans, Nederlands, Pools, Zweeds, Spaans
Display
LC-display
Electrical safety
CE, cCSUit
Max flow
Zonder geïntegreerde meetgaspomp: 5 l/h ... 100 l/h
Met geïntegreerde meetgaspomp: 30 l/h ... 60 l/h
Met geïntegreerde meetgaspomp: 30 l/h ... 60 l/h
Downloads
Brochures
Manuals
Certificates
-
CB Test certificate S700
-
EC-Type Examination certificate 4
-
EC-Type examination certificate 6
-
Product Conformity certificate UNOR MUTOR OXOR P OXOR E
-
Statement of Conformity 1
-
Statement of Conformity
-
TUV QAL1 S700 MULTOR module
-
TUV QAL1 S700 UNOR module
-
Certificate of Compliance
-
EC-Type examination certificate 5
-
EC-Type examination certificate 7
-
Statement for EC Type Examination Certificate TV97ATEX1207X
-
Statement of Conformity 2
-
QAL3 beschrijving
-
TUV QAL1 S700 OXOR-P module
Features:
- 5 different measurement principles available
- More than 60 measuring components to choose from
- 3 different housing variants for different fields of application
- Up to 3 analyzer modules in a single housing
Benefits:
- Easy application-specific adaptation due to modular design
- Also suitable for zone 1 and zone 2 explosive environments (ATEX)
- Automatic adjustment with test gas or calibration cell
- Integrated self-monitoring and fault diagnosis
Applications:
- Emission measurement according to 13th (2001/80/EC), 17th (2000/76/EC), and 27th German Federal Emission Protection Directive (BImSchV).
- Cooling gas monitoring of turbo generators
- Measurement of CO for coal mill monitoring
- Purity measurement of H2 in pressure swing adsorption plants
- Monitoring of CO2 in natural gas in natural gas conditioning plants