Degassed Cation Conductivity

Degassed Cation Conductivity Manvia

 

After the sample has been brought to the boil and all the gas has dissolved, it is cooled to 25-35 °C, and then the treated sample is placed in a conductivity measuring cell. Atmospheric pressure compensation of the boiling point is performed automatically to ensure correct sample degassing with minimal energy waste.

Degassed cation conductivity is an important parameter for the boiler and turbine. Most turbine manufacturers have specified that the value of cation conductivity in the steam must be below a certain conductivity limit (less than 0.5 μS/cm) to prevent damage to the turbine.

Manvia, our supplier, has various solutions in the field of water panels or Swas panels. Special analyzers such as Silica, Hydrazine, Sodium and Dissolved Oxygen can also be added here

 

Kijk voor analyzers op onze ander pagina’s zoals geleidbaarheid en pH meting

 

 

Specifications

System
flexibel
Pressure / temperature
4 bar(g)[58 psi(g)] / 50ºC (323ºK)
IP Raiting
IP66
Housing
SS 316
sample volume
80 - 250 ml /min.
Dimensionsd (H x W x D)
900 x 500 x 315 mm
Display
manvia analyser
Weight
Depanding on configuration
Analyse frequecy
continue

Video

Characteristics of degassed cation conductivity

  • According with ASTM D4519
  • Provide with necessary shell protection to prevent operators from hot injury
  • Compact wall mount design
  • CE marking
  • Flow indication
  • Switch protection “no flow”

Applications

  • Degassed measurement in Steam Water Analysis Systems

Benefits of the Manvia degassed cation conductivity

  • Include atmospheric pressure sensor to set the temperature at boiler point
  • Designed to install the cationic resin column, the transmitter and the conductivity probes in the same plate
  • Short time to reach the degassed conductivity measurement
  • Including all necessary components as a complete unit, just piping, plug-in and ready to use